🇺🇸 FDA

QA/QC Tools · OOS · FDA Guidance · GMP · ICH Q10

OOS Investigation Template

Structured Out-of-Specification investigation following FDA's 2006 OOS Guidance. Complete Phase I laboratory investigation before proceeding to Phase II full investigation. Auto-saves locally. Print as PDF.

1
Phase I — Lab Investigation
2
Phase II — Full Investigation
0% complete
Section 1 — OOS Record Header
Section 2 — Phase I: Laboratory Error Checklist

Complete all items below. "Problem Found" means a laboratory error was identified. "No Problem Found" means this item is not the cause of the OOS result.

1 Was the correct test method / procedure used?
2 Was the reference standard within its expiry date and properly stored?
3 Was the instrument calibration current and within qualification period?
4 Was the sample properly prepared per the test method instructions?
5 Were all calculations reviewed and found to be free of errors?
6 Was instrument performance verified prior to and during analysis (e.g., SST)?
7 Were there any transcription, data entry, or data transfer errors identified?
8 Was the analysis environment suitable (temperature, humidity, vibration, etc.)?
9 Did the HPLC / instrument system suitability test pass per method criteria?
10 Was the column (or other consumable) used within its qualified lifespan?
Section 3 — Phase I Conclusion
Section 4 — Phase II: Full OOS Investigation

Extend the investigation beyond the laboratory. Review manufacturing, sampling, and authorize additional testing if appropriate.

Manufacturing Investigation
Was the correct batch record formula used?
Were correct raw materials (identity, quantity, certificate of analysis) used?
Was the correct equipment used and was it qualified / clean?
Were all critical process parameters within their specified ranges?
Sampling Investigation
Was sampling performed per the approved sampling procedure / plan?
Was the sample quantity adequate for the test?
Were sample storage conditions (temperature, light, humidity) correct?
Additional Testing
Section 5 — Disposition
Section 6 — Approvals

Frequently Asked Questions

An OOS (Out-of-Specification) result is any test result that falls outside the established acceptance criteria defined in the product specification, registration dossier, or official compendia (e.g., USP, EP). OOS results require formal investigation per FDA guidance and EU GMP. The investigation must determine whether the result represents a real product quality failure or is attributable to an identifiable laboratory error.
The FDA issued "Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production" guidance in October 2006 (finalizing a 1998 draft). It describes a two-phase structured investigation: Phase I (laboratory investigation to identify assignable laboratory error) and Phase II (full investigation including manufacturing review, sampling assessment, and additional testing). The guidance explicitly prohibits discarding OOS results without scientific justification and sets strict limits on retesting.
OOS (Out-of-Specification) results exceed or fall below the defined acceptance limits. OOT (Out-of-Trend) results are within specification but show a statistically unexpected trend over time, detected through statistical process control or stability trending. OOT results indicate a potential developing quality problem. Both require investigation, but OOT investigations are proactive (before failure) while OOS investigations are reactive (after failure).
Yes, but only under strictly controlled conditions. An OOS result may be invalidated if — and only if — a specific, documented, assignable laboratory error is identified (e.g., wrong reference standard used, confirmed calculation error, documented instrument malfunction). Statistical outlier tests and lab retesting without identified cause are not acceptable for invalidation under FDA guidance. The original OOS result must always be retained in the analytical record.
Phase II is required whenever Phase I laboratory investigation concludes with no assignable laboratory error. Phase II extends the investigation to include: manufacturing records and process review, raw material verification, sampling procedure assessment, environmental monitoring data, and authorized additional testing. Phase II must be pre-planned, with the number of additional samples and testing criteria defined before testing begins.